ATEC

Products
+
  • 1(2).png
  • 2(1).png
  • 01.png
  • 02.png
  • 03.png

High and low temperature damp heat test chamber

In order to pursue better quality and more stringent reliability guarantee, ATEC has independently developed the whole series of the test chamber and formed standardization since 2006. Continuous improvement and absorption of advanced technology have formed "ch" and "Su" series to meet the needs of different customers.

Series:

SU Series

Temperature Range::

150 ℃ ~ -40 ℃ or -70 ℃

Cooling Rate:

1 ℃/min or 3 ℃/min

Details

1

In order to pursue better quality and more stringent reliability guarantee, ATEC has independently developed the whole series of the test chamber and formed standardization since 2006. Continuous improvement and absorption of advanced technology have formed "ch" and "Su" series to meet the needs of different customers.

This series of products are equipped with volume models such as 250L, 400L, 600L to 2000L. In terms of design, environmental protection refrigerant and high-performance refrigeration system are adopted. The main parts are international well-known brands and unique humidification system, providing a variety of humidity range options.

Temperature range: 150 ℃ ~ - 40 ℃

 

150℃~ -70℃

 

Cooling rate: 1 ℃ / min or 3 ℃ / min

 

Model description

 

Ch: Product Series No. 250: Studio volume (unit: l)

 

T: Single temperature, without humidity, no "t" indicates with humidity

 

C: Indicates the minimum temperature

 

C: Minimum - 70 ℃; None: minimum - 40 ℃; E: Minimum - 20 ℃

 

Technical parameter

型号(SU)100(C)160(C)250(C)400(C)600(C)1000(C)1600(C)2000(C)
Nominal internal volume (L)100L160L250L400L600L1000L1600L2000L
Performance
Temperature range(T/TC)(-40℃/-60℃)~+120℃-40℃~+150℃/ -70℃~ +150℃
Humidity range

Ch series: 20% ~ 98% RH (humidity working range + 20 ℃ ~ + 85 ℃)

Su series: 10% ~ 98% RH (humidity working range + 20 ℃ ~ + 90 ℃)

Temperature fluctuation±0.3℃
Temperature deviation±2℃
Humidity deviation-3 ~ + 2% RH (above 75% RH), ± 5% RH (below 75% RH)
Heating rate (whole process average)CH series: 2 ℃ / min Su series: 3 ℃ / min
Cooling rate (whole process average)CH series: 1 ℃ / min Su series: 2 ℃ / min
RemarksThe above indexes are measured under no-load and ambient temperature of 20 ± 2 ℃, in which the temperature rise and fall rate is measured at ± 85 / - 30 ℃ and + 85 / - 55 ℃
Internal dimensions(mm)
D400500580700750100012501400
W500560600700850100010001200
H500600740850900100012001200
External dimension (mm)
D12601530165517251850208022502450
W7508108509501100125012501450
H(Including casters)17651840182018852000210523002300
Casters4 standard casters
Observation window450x450 with heating device to prevent condensation and frosting
Test hole1 × 50, right side of box1 × 80, right side of box2 × 100, one on the left and right sides of the box
Stainless steel shelf (2 pieces as standard)Bearing capacity: 25kg / pieceBearing capacity: 50kg / piece
lighting35W/12V2*35W/12V
Cooling modeAir cooling
Power Supply380V±10%/50HZ±1,Three phase five wire system (3P + N + G)
Installed capacity(kw)(T/TC)3.6/4.44.5/5.65.2/7.46.0/8.58.0/119.0/1411/1613/17
Meet the test methodGB2423.1、GB2423.2、GJB150.3、GJB150.4、GB11158、GB10589
Safety protection measuresPower supply undervoltage, overvoltage, phase loss and phase sequence error protection; Triple overtemperature protection; Heater short circuit protection; Refrigerator overpressure / overload protection; Overload / overheating protection of blower motor; Control board overcurrent protection; Leakage protection.
OptionalRecorder, special test hole, special shelf, timing function, water cooling mode, Internet function, etc
 

Other Products

High and low temperature damp heat test chamber

CH Series

Temperature range:

150 ℃ ~ -40 ℃ or -70 ℃

Cooling rate:

1 ℃/min or 3 ℃/min

High and low temperature damp heat test chamber

SU Series

Temperature range:

150 ℃ ~ -40 ℃ or -70 ℃

Cooling rate:

1 ℃/min or 3 ℃/min

Rapid temperature change test chamber

Temperature range:

-70 ° ℃ ~ 150 ° ℃

Cooling rate:

5 ℃ ~ 25 ℃ / min (mechanical refrigeration)

Temperature shock test chamber

Hatchback Basket

Temperature range:

-65 ℃ ~ 150 ℃

Cooling rate:

H type is about 4 ℃ / min, M type is about 3 ℃ / min, L type is about 2 ℃ / min (- no-load average value in the range of 55 ° C

Temperature shock test chamber2

Three-compartment

Temperature range:

-65 ℃ ~ 150 ℃

Cooling rate:

H type is about 4 ℃ / min, M type is about 3 ℃ / min, L type is about 2 ℃ / min (- no-load average value in the range of 55 ° C

Temperature and humidity vibration three comprehensive test chamber

Three comprehensive test box

Temperature range:

-70 ℃ ~ 150 ℃

Cooling rate:

1 ℃ - 5 ℃ / min optional (no load, whole process average. - 40 - + 100 ℃)

I want to consult